+91 70565 55582[email protected]Get a quoteRequest a call
RTS Calibration
Menu

Holiday detector testing

A coating holiday detector searches for discontinuities in a protective coating. A low-voltage wet-sponge instrument and a high-voltage type are checked differently. A function check for an alarm and a measurement of the selected output are also separate results.

Send an enquiry
Holiday detector — equipment illustration
Equipment illustration
On this page

Continuous and pulsed detector outputs

Record the model, applicable method, range selected and the accessory used to verify output. Some instruments have a pulsed output, so the measuring system and reported quantity need to be appropriate to that design. Include the return lead and electrode if assembly checks are required.

Holiday detector output and accessories

List the output setting, electrode, cable and return connection specified for the job. RTS checks the specified output and functional indications in the identified electrical arrangement. An output check does not establish coating integrity on a structure, and instrument work must not be substituted for the specified inspection.

Specifications and service checks

An alarm on a deliberately exposed area confirms a response under that setup. It does not give a measured output voltage. Distinguish the coating inspection check from a quantitative output comparison, and identify the applied-voltage mode.

Range and configuration
A holiday detector’s working value is the test voltage applied to a coating. A high-voltage DC detector and a low-voltage wet-sponge instrument use different operating ranges.
Manufacturers and models
Elcometer 266 is a high-voltage DC holiday detector. Identify the probe handle as well as the base instrument.
What we check
Have the test output checked at the working settings using equipment rated for that waveform. For an alarm check, the test arrangement should represent a discontinuity in the coating under inspection.
Reference equipment and accuracy
A calibrated high-voltage divider or suitable dedicated verifier must be rated for the output and have adequate input impedance. Specify whether the voltage is continuous or pulsed.
When to repeat the check
Check the return lead and alarm function before work. Repeat output verification after service to the probe handle, cable or high-voltage electronics.

Technical references Elcometer 266 Holiday Detector

Enquire about Holiday detector

Service enquiry