Hot and cold chamber qualification

A chamber that alternates between hot and cold needs observations on both sides of the cycle. Two steady readings cannot establish the time taken to move between them. Identify the hold periods and direction of travel required in the test.

Send an enquiry
Hot cold chamber — equipment illustration
Equipment illustration
On this page

Timing the heating and cooling transitions

If a controller starts its timer before every monitored point has settled, record when the programmed hold starts and when the working volume reaches the required band. The difference can affect how long the item actually experiences the specified condition.

A test sheet may quote a transition as ‘20 °C to −10 °C in 15 minutes’. For that requirement, the start and finish temperatures must refer to a named location, such as the chamber air near the item or a probe attached to it. Record which direction of travel the limit applies to. Cooling and warming can give different times.

Specifications and service checks

A chamber cycling from cold to hot may show a setpoint step long before a shelf probe reaches the new condition. If transition time matters, identify the start and finish temperature thresholds and the probe location. Those details determine which interval is read from the trace.

Range and configuration
The ESPEC SU-242 is a temperature-only example rated from −40 to +150 °C. Useful volume, loading and the points named on the order determine the comparison.
Manufacturers and models
The SU and SH models in ESPEC’s benchtop series include different temperature and humidity configurations.
What we check
Map selected positions in the working volume at the agreed settings. Identify where the control sensor sits and report the difference between its indication and the reference readings.
Reference equipment and accuracy
Use a set of calibrated temperature sensors and a synchronised logger. The reference uncertainty should support the allowed variation within the test space.
When to repeat the check
Review after refrigeration or heater repair, a change of control probe or a significant change in loading. Compare repeat maps at the same locations.

Technical references ESPEC SU and SH benchtop chamber specifications · EURAMET climatic chamber calibration guide