Humidity chamber qualification

Relative humidity performance in a chamber is tied to temperature. Request each humidity point as a temperature and humidity pair, with the rack positions and load condition identified. A single value at the built-in probe cannot show differences across the shelf.

Send an enquiry
Humidity chamber — equipment illustration
Equipment illustration
On this page

Humidity variation during the hold

If your procedure specifies a holding period, identify the portion used to assess variation. Keep the controller indication and independent measurements in separate columns. That format helps distinguish an indication error from a difference between locations.

A chamber can hold its centre sensor at the chosen humidity while one occupied shelf runs warmer. Because relative humidity also depends on temperature, a paired temperature and humidity trace at that shelf is more useful than comparing its humidity value with the centre reading alone.

Specifications and service checks

A shelf hygrometer placed close to a chamber outlet may read differently from one beside the product. If the question is about the product position, identify that position and the associated temperature setpoint. Relative humidity comparisons need both position and temperature to be meaningful.

Range and configuration
BINDER quotes 10–80% RH for the KBF LQC 240. Check the operating diagram for the particular combination of temperature and humidity needed in your workspace.
Manufacturers and models
BINDER’s KBF LQC 240 is one example of a constant-climate chamber. Its specified envelope belongs to that model.
What we check
For a specified relative humidity, identify the corresponding chamber temperature. Compare workspace humidity and temperature together, documenting the shelf locations of the reference sensors.
Reference equipment and accuracy
Use calibrated temperature and humidity probes with corrections valid at the requested conditions. The uncertainty must allow for spatial variation and sensor response.
When to repeat the check
Review after relocation, a sensor change or work on the humidifier. A check near the routine storage condition can reveal drift between mappings.

Technical references BINDER KBF LQC 240 data sheet · EURAMET climatic chamber calibration guide