Reference bath or automatic ice-point device
Identify whether the unit is a maintained ice bath or an automatic reference device. If several probes share the bath, state the typical occupancy and which positions serve as references. This separates a local comparison from an assessment of the entire working space.
Ice point baths as reference arrangements
Identify the vessel, working depth, probe load and the check your laboratory requires. Separate verification of the realised reference condition from calibration of a thermometer placed inside it. RTS verifies the realised ice-point condition with the preparation, immersion and probe loading recorded.
Specifications and service checks
With several probe wells, one may be used every day and another only for a spare. Identify the well number and the usual depth of the sensing element. Comparing both positions gives a better answer to a well-to-well difference than repeating one well under a different label.
- Range and configuration
- An ice-point reference realises a single temperature near 0 °C. It does not establish the full-span performance of a sensor immersed in it.
- Manufacturers and models
- Isotech documents automatic ice-point references for temperature and reference-junction work.
- What we check
- Check the realised temperature and stability in the working zone. For a multiwell reference, identify the wells and loading used.
- Reference equipment and accuracy
- A calibrated reference thermometer with suitable low uncertainty at 0 °C is needed. The freezing point reference and the triple point of water are distinct reference conditions.
- When to repeat the check
- Verify a freshly prepared reference when used. A maintained automatic reference also needs periodic checks and review after service.