Infant incubator testing

The temperature shown by an infant incubator is only one reading in an equipment performance check. Air temperature at different locations, an attached temperature probe and the controller can produce three different sets of results.

Send an enquiry
Infant incubator — equipment illustration
Equipment illustration
On this page

Probe positions in the cabinet

Use the mattress as a reference when labelling measurement positions. A drawing marked with the door side and head end is easier to reproduce than five labels reading ‘corner’. Identify the air control or other mode defined by your engineering procedure and any separate monitoring probe. If humidity is included, give the working setting and the location of the comparison sensor.

Infant incubator engineering assessment

A temperature indication, chamber distribution, humidity display and alarm record may need separate treatment. Include the required conditions and specialist test-system details for review. This enquiry concerns equipment released for bench assessment by biomedical engineering, not instructions for changing conditions around an infant.

Specifications and service checks

After a mattress tray or canopy change, an old probe-position sketch may describe a different cabinet arrangement. Mark the current tray height and probe positions on the bench plan so repeat temperature observations refer to the same locations.

Range and configuration
Air indication, a skin-probe input and chamber temperature distribution describe different quantities. Record the option and probe identity for each requested result.
Manufacturers and models
Dräger lists the Isolette 8000 with air and skin temperature measuring functions.
What we check
A service assessment can compare indications and map chamber positions under the manufacturer’s engineering test conditions. Identify mattress configuration when it affects probe positioning.
Reference equipment and accuracy
Use calibrated temperature probes or an incubator analyser with known channel uncertainties. Humidity or air velocity needs a reference for that quantity if included.
When to repeat the check
Repeat the affected checks following sensor, controller or airflow work. Use the facility’s acceptance and periodic safety programme for scheduling.