On this page
Which extension rods set the bore size?
List the rods or spacers actually combined for the working diameters. If the head is set against a reference before use, identify that reference. Keep the result with the measuring head and rod combination used for each point.
A large bore may require two extensions on the same head. Recording only the head serial number leaves no way to tell which assembly was checked if another extension of the same nominal size is fitted later.
Specifications and service checks
When an extension is changed, a speck on a joint face affects the whole assembled length. If a disagreement occurs only with one extension fitted, identify the head, that extension and the setting reference at the corresponding length.
- Range and configuration
- Mitutoyo’s Series 137 catalogue lists a 50–63 mm head with 13 mm travel and sets that extend the span with rods. A head-only result therefore describes only one configuration.
- Manufacturers and models
- Mitutoyo Series 137 tubular inside micrometers use a measuring head, anvils and extension rods.
- What we check
- Compare each ordered combination with known internal lengths, using the working contacts and rods. Include different screw positions and mark the rod combination for each result.
- Reference equipment and accuracy
- Use calibrated setting rings or an internal-length comparison arrangement with appropriate gauge blocks and accessories. Support longer assemblies to control sag and alignment error.
- When to repeat the check
- Keep the head and rods identified as a set. After replacing an extension or damaging its thread, compare the affected combinations before using them for acceptance work.
Technical references Mitutoyo Tubular Inside Micrometers · Mechanical Tubular Inside Micrometers Extension Rod Type - Mitutoyo