Test current and the reference resistor
Identify the working range, reference resistor, normal test current and terminal arrangement. If a separate null detector is used, list it as part of the setup. State whether the bridge ratio, direct indication or an assembled low-resistance measurement needs evaluation.
If an unfamiliar low-resistance value appears only after moving the sense clips, note their exact positions. A sense connection on the wrong side of a joint can include its drop in the measured resistance even when the bridge ratio is correct.
Specifications and service checks
A bridge comparison can reverse current direction to help recognise a thermal voltage in the connection. If the observed balance changes with polarity, tell the bench both readings. A single average can hide the clue that distinguishes the offset from the resistance.
- Range and configuration
- One milliohm is 1,000 microohms. Provide both the working resistance and the test current, since self-heating can change a low-ohm standard.
- Manufacturers and models
- Megger DLRO10 is a digital four-terminal low-resistance ohmmeter. A traditional Kelvin bridge uses a ratio comparison for similar low-resistance work.
- What we check
- Compare balance points for four-terminal resistors at known ratios. Record measuring current, ratio arm position and lead arrangement, and compare forward and reversed current readings when thermal effects are relevant.
- Reference equipment and accuracy
- Use low-resistance standards rated for the applied current, with separate current and potential connections. Include thermal voltage and self-heating in uncertainty.
- When to repeat the check
- Trend a stable low-ohm checkpoint. Recheck after lead, current-source or selector repairs.
Technical references DLRO10 and DLRO10X digital low resistance micro-ohmmeters