Voltmeter and divider identification
Give the assembly serial numbers, AC or DC mode, frequency or polarity, and the working range. Specify whether a single system indication or separate divider ratio and readout results are required. Use the configuration prescribed for the instrument.
A system can display kilovolts while the receiving meter measures only volts. If a replacement readout is installed, confirm both its input loading and the ratio applied in software. A correctly working divider does not by itself establish the new indicated kilovolts.
Specifications and service checks
A display may report peak voltage on one range and RMS on another. On a non-sinusoidal source, a simple fixed conversion between the two can be misleading. Give the waveform and indicated quantity used in the test instruction.
- Range and configuration
- A 1,000:1 divider produces a nominal 1 V per 1 kV at its design load. Its AC response and permissible voltage depend on the individual rating.
- Manufacturers and models
- Ross Engineering supplies high-voltage measuring systems and matched probe/meter pairs. Record the divider ratio and the readout input specified for the assembly.
- What we check
- Compare indicated high voltage or divider ratio with an independently measured voltage at the requested polarities and frequencies.
- Reference equipment and accuracy
- Use a calibrated high-voltage reference divider and suitable measurement system. Loading, leakage and frequency response affect the ratio uncertainty.
- When to repeat the check
- Recheck the pair after readout replacement, damage to insulation or repair to the divider.