Kinematic viscometer calibration

A gravity capillary viscometer relates a measured flow time to kinematic viscosity using the constant assigned to the specific glass tube. The bath temperature matters because liquid viscosity can change appreciably with temperature.

Send an enquiry
Kinematic viscometer — equipment illustration
Equipment illustration
On this page

Tube constant and working temperature

List each tube number, its marked constant, the working temperature and the liquid range used. Identify whether the request covers the tube constant, timing device, bath or assembled measurement. Replacement tubes should retain their own result even when used in the same holder.

Two capillary tubes may be held in the same bath at the same temperature but carry different constants. Keep each flow time with the number etched on its tube. Averaging the times before applying their individual tube constants can produce a result that belongs to neither instrument.

Specifications and service checks

A capillary with a working constant of 0.10 mm²/s² and a corrected flow time of 100 s corresponds to 10 mm²/s when the method uses the simple constant-times-time relation. Use the assigned constant and any corrections required for the actual tube.

Range and configuration
Report kinematic viscosity in mm²/s and the bath temperature. For a capillary, the measured flow time and the calibrated constant determine the result, with corrections where applicable.
Manufacturers and models
CANNON SimpleVIS uses gravity flow through a capillary. Manual capillary instruments also carry an individual tube constant.
What we check
Compare measurements of certified viscosity oils at the specified bath temperature. On a manual tube, establish or verify its constant over the suitable flow-time band.
Reference equipment and accuracy
Use certified kinematic viscosity standards, an independent calibrated thermometer and a traceable timer. Temperature uniformity can dominate the result.
When to repeat the check
Check reference oil performance after cleaning or moving a tube. A persistent change in flow time warrants review of the constant.

Technical references SimpleVIS Portable Automated Kinematic Viscometer