Working masses including the tare vessel
State the balance capacity, displayed division and actual masses handled, including containers. If the instrument is used inside an enclosure, describe the bench and the airflow around it. The calibration result should identify the location when the balance is checked in its usual installed condition.
For a before-and-after mass loss determination, the change may be small compared with the gross mass of the container. Supply the approximate change as well as the gross load when selecting points for the instrument.
Specifications and service checks
For a filter weighed before and after use, a container ID ties the two observations together. If the gross readings are recorded to 0.001 g, keep those digits through subtraction and round the reported mass change at the end.
- Range and configuration
- An OHAUS PR2202, for example, weighs to 2,200 g with a 0.01 g display division. The smallest sample a process can accept depends on measured uncertainty.
- Manufacturers and models
- OHAUS PR2202 is a precision-balance example. A different suffix can change capacity or readability.
- What we check
- Compare the balance across its useful load range and repeat weighings at a representative net sample mass. If a draft shield is fitted, use the same operating position as the laboratory.
- Reference equipment and accuracy
- Choose weight class and certificate uncertainty against the smallest permitted weighing error. A class marking does not replace the individual weight certificate.
- When to repeat the check
- Set the recall using check-weight performance at both the light and heavy end. Review it after a change of bench or weighing range.