On this page
Bath temperature, carrier speed and run time
A bath probe may reach the set temperature before liquid inside a closed test vessel. If the method specifies vessel temperature, mark the vessel and where the comparison takes place. For repeat work, give the carrier position and normal loading. We can record that temperature measurement alongside the carrier speed and cycle time without confusing a bath reading with a vessel reading.
Launderometer temperature and rotation
Identify whether the requested temperature concerns the bath, a vessel or another defined location. Keep the container and accessory set with the enquiry. A reading at one bath position cannot automatically represent every vessel or establish the washing result of a fabric specimen.
Specifications and service checks
If a vessel has been moved to another carrier position, retain its identifier and station in the comparison record. A bath temperature reading at one point and the motion of the loaded carrier answer separate questions.
- Range and configuration
- The published rotation rate for the M228 family is 40 ±2 rpm. Temperature settings and vessel sizes also belong on the equipment record.
- Manufacturers and models
- The SDL Atlas M228 Launder-Ometer is a rotating-vessel wash fastness tester.
- What we check
- Compare the bath indicator with an independent probe. Measure rotational speed separately and time a complete programme at a setting the laboratory uses.
- Reference equipment and accuracy
- A calibrated immersion thermometer, optical tachometer and timer cover three different quantities. Choose the probe uncertainty for the permitted bath variation.
- When to repeat the check
- Check the bath after a heater or controller change and speed after drive work. Check the timer when a programme or controller is replaced.
Technical references SDL Atlas product reference