Level sensor calibration

Level sensors respond to different physical quantities, including distance to a surface, pressure head or float position. The raw sensor result and the level value calculated by the plant do not always use the same reference point.

Send an enquiry
Level sensor — equipment illustration
Equipment illustration
On this page

The sensor datum and the physical level

Identify the sensing principle and the measurement datum. For a distance sensor, include the mounting position and empty reference distance. For a pressure-based sensor, give the liquid density used. Those details determine how a reference level should be compared with the signal.

For a top-mounted distance sensor 2.5 m above the tank floor, a 0.5 m reading from the sensor to the surface corresponds to a level of 2.0 m above the floor. If the sensor and the operator’s screen show those two values at the same time, both can be right.

Level sensors and measurement principles

Include the reference datum and whether the result is distance to a surface, filled height or another quantity. RTS calibrates the sensor using a physical reference matched to its measurement principle. If only the signal electronics will be supplied, make that boundary clear rather than treating the absent sensing assembly as included.

Specifications and service checks

A radar or ultrasonic sensor can report distance to the surface while the controller displays depth of product. If the trend moves down as the tank fills, identify whether the raw distance or the converted level is being logged before comparing the sensor.

Range and configuration
A sensor can report distance down from its face or level up from the vessel datum. The same target movement changes these readings in opposite directions.
Manufacturers and models
VEGA lists measuring distances up to 120 m for VEGAPULS 6X options. Actual range depends on the selected antenna, version and application. Record the instrument reference plane and configured blanking distance.
What we check
Compare indicated distance at known target positions, or compare an installed level reading with a traceable liquid level where appropriate.
Reference equipment and accuracy
Use a calibrated length or level reference and a target adequate for the sensing principle. Alignment and the reference plane affect uncertainty.
When to repeat the check
Review after relocation, a new tank reference or changes to stored interference echoes.

Technical references Data sheet - VEGAPULS 6X - Radar sensor for continuous level measurement of liquids and bulk solids