Choosing the lamp and sample positions
A fabric swatch can match under one lamp and differ under another. Which lamp names appear on the comparison sheet? Tell us whether staff use the middle of the booth or lay several samples across it. A level reading at the sample plane, a map of the light across that plane and a measurement of spectral output each answer a different question. Lamp age and replacements help when you are comparing visits.
Light booths used for visual comparison
Record the lamp identities, selected lighting modes, working area and positions where measurements are required. Say whether the enquiry concerns illumination level, colour characteristics or a wider assessment of the viewing conditions. RTS assesses the booth at the specified viewing positions and lighting modes, with illumination and colour-related results identified separately.
Specifications and service checks
For a colour approval using several illuminants, note the selected lamp mode and viewing position where the mismatch appears. If the comparison changes after a lamp replacement, the source identity and operating history give more context than the booth cabinet serial alone.
- Range and configuration
- The measurement may concern illuminance in lux, spatial uniformity, spectral distribution or chromaticity. Equal lux readings from two lamps do not establish a colour match.
- Manufacturers and models
- X-Rite SpectraLight QC provides selectable viewing illuminants. Identify the illuminant setting, working plane and booth configuration used for assessment.
- What we check
- Measure a defined grid on the viewing plane, then evaluate the spectral or colour characteristics at the specified locations and settings.
- Reference equipment and accuracy
- Use a calibrated illuminance meter for level and a suitable spectroradiometer for spectral results. Record instrument geometry and ambient conditions.
- When to repeat the check
- Trend output against lamp hours and verify again after relamping or changes to booth interior surfaces.