Liquid nitrogen comparator calibration

A comparator for low-temperature probes has a working region where the reference and test probe are compared. The reported temperature comes from the reference and the comparison conditions. The name of the coolant on the vessel does not, by itself, give the temperature reached by each probe. Tell us how the sensors are held and which reading your procedure uses.

Send an enquiry
Liquid nitrogen comparator — equipment illustration
Equipment illustration
On this page

Probe position in the comparison region

For a comparison against a reference probe, give the probe diameters and the immersion depth the holder permits. The comparison needs the reference and the unit under test at identified positions and times. If your method uses a boiling point as its reference instead, send the pressure record that goes with the observation. The two approaches have different reference information.

Liquid nitrogen comparators

State whether the request is for comparison of customer probes or evaluation of the comparator itself. Cryogenic work needs a specific capability and method review. RTS evaluates the comparator or compares the customer probes under the specified cryogenic conditions, with the reference sensor and pressure recorded.

Specifications and service checks

Two comparison positions can place sensing elements at different depths even if their probe stems are equal in length. Mark the sensing length and normal insertion for each probe. Keep the measured reference temperature with the run rather than assigning a temperature solely from the name of the bath.

Range and configuration
A liquid-nitrogen comparator provides a nominal point around −196 °C. Its actual realised temperature depends in part on pressure. A controlled cryostat has a different specification.
Manufacturers and models
Isotech lists liquid-nitrogen comparator equipment and distinguishes it from the Model 459 controlled cryostat.
What we check
Compare the working position with a traceable low-temperature reference and state the conditions on the certificate. A nominal boiling point alone cannot supply a correction.
Reference equipment and accuracy
Use a reference suitable for cryogenic temperatures and a calibrated readout. Include any pressure correction and depth or position effects in the uncertainty evaluation.
When to repeat the check
Check the reference after relocation or a change in probe holder. The transfer probe’s calibration interval is a separate consideration.

Technical references Isotech laboratory temperature calibration catalogue