On this page
Stability, flow and travel rate
An older Marshall frame may use a proving ring with its own conversion chart. If so, send the ring identity and the chart it is actually used with. A newer frame may display force directly from a load cell. In either case, give the normal load range, the flow indicator units, the breaking head fitted and the carriage speed. Force and flow need separate results on the report.
Marshall stability apparatus
The loading head, working range and specimen-conditioning equipment are identified with the test arrangement. Each channel has its own reference and result, allowing the laboratory to trace both the applied load and deformation used in its material-test calculation.
Specifications and service checks
If an old Marshall record lists both peak load and flow at peak, the force and displacement channels must be identifiable at the same event. Tell the workshop whether your system records them together or relies on separate indicators.
- Range and configuration
- The B042-KIT platen rate is 50.8 mm/min. Force capacity, flow travel and rate each need their own units and acceptance limit.
- Manufacturers and models
- Matest B042-KIT pairs a Marshall load frame with a 30 kN proving ring.
- What we check
- Verify the force channel across the requested loads, the flow indicator against known displacements and platen speed over a timed travel.
- Reference equipment and accuracy
- Use a calibrated force-proving standard, a displacement reference and a timer. The proving ring needs a valid certificate for its loading direction.
- When to repeat the check
- A new proving ring changes the force chain. Review the next due date against frame use and its previous as-found errors.
Technical references Matest product reference