Microcentrifuge calibration

Compact microcentrifuges may use one fixed rotor or interchangeable small rotors. Some display speed numerically, while others expose only programme settings. Identify the rotor marking and which quantity the instrument actually indicates before ordering a comparison.

Send an enquiry
Microcentrifuge — equipment illustration
Equipment illustration
On this page

What the timer includes in the run

If the timer includes acceleration or braking in the displayed duration, note the event that starts and ends the interval in your equipment procedure. A measured running time is hard to interpret without that definition.

If the laboratory times a run from pressing start while the instrument starts its countdown only after reaching speed, the displayed interval and the operator’s stopwatch will disagree. Define the two timing events first. Their difference is not automatically an error in the timer.

Specifications and service checks

A microcentrifuge may hold one size of tube directly and a smaller size through an adapter. If a routine calculation uses centrifugal force, identify the adapter and where the tube sits in the rotor. Its position affects the radius used to relate rpm to force.

Range and configuration
Eppendorf’s MiniSpin plus has a 14,500 rpm maximum. Beckman Coulter lists 100,000 rpm for the Optima XPN-100 ultracentrifuge. Always use the limit of the installed rotor and drive combination.
Manufacturers and models
Eppendorf identifies separate MiniSpin and MiniSpin plus versions. Record the instrument’s precise version and fitted microtube rotor on the order.
What we check
For instruments with indicated speed, request a comparison using the manufacturer-approved service method. A fixed-speed mini unit can be reported at its operating speed.
Reference equipment and accuracy
Speed traceability must suit the range and the drive’s authorised measurement arrangement. A regular handheld tachometer may not suit a sealed high-speed system.
When to repeat the check
Check the speed indication after drive or sensor service. Keep the rotor service-life record separate from the calibration date of the speed measurement.

Technical references Eppendorf MiniSpin documentation · Beckman Coulter Optima XPN-100