Matching capsule sensitivity to the readout
A replacement capsule can fit an existing preamplifier yet have a different sensitivity. When a field instrument uses a sensitivity entered from a certificate, tell us which capsule that value belongs to and which unit the display expects. Include the microphone model, frequency range, grille or adapter, acoustic field correction if used, and the preamplifier serial number if the pair is to be calibrated together.
Measurement microphones
Specify whether the enquiry concerns sensitivity at selected frequencies or a broader response. Keep the pressure or sound-field convention from the previous report. RTS measures the microphone assembly’s acoustic response; an electrical display check is recorded as a separate function.
Specifications and service checks
A bare capsule sent from an environmental monitor and a complete microphone with preamplifier represent different measuring assemblies. If the quantity of interest is the voltage at the recorder input, identify the preamplifier and any conditioning gain in that path.
- Range and configuration
- Sensitivity is commonly stated in mV/Pa at a reference frequency. A pressure-field or free-field frequency response is a separate specification.
- Manufacturers and models
- GRAS 40AE is a half-inch free-field measurement microphone. Identify the capsule and preamplifier as a pair where both are used.
- What we check
- Compare sensitivity at a reference sound pressure, or request response across a stated frequency band and field type.
- Reference equipment and accuracy
- Use a calibrated acoustic coupler or reference microphone system suitable for the capsule. The reference pressure, coupler correction and environment influence uncertainty.
- When to repeat the check
- A field calibrator provides an intermediate check. Review the capsule after a drop, moisture exposure or a change in sensitivity.
Technical references GRAS 40AE measurement microphone data