Scale factors for the objective and camera
A measurement made with the 10× objective has its own scale factor. Changing to 40×, fitting a different camera adapter or changing the camera resolution can change how much specimen length each pixel represents. For a measurement check, list the objective and the saved image setting your inspector actually uses. A captured image of a reference scale can then be traced to the correct configuration.
Microscope scale and image measurement
RTS checks the specified measurement configuration with the objective, camera and software settings identified. Each magnification or camera arrangement used for dimensional work needs the applicable scale recorded. The reference artefact and laboratory procedure remain with the results so measurements can be reproduced after a setting or component change.
Specifications and service checks
A stage micrometer viewed through a 10× objective establishes an image scale for that configuration. Switching objectives or changing camera binning can alter how many pixels represent the same distance. Identify each objective and camera setting used for actual dimensional measurements.
- Range and configuration
- Optical magnification, digital zoom and spatial resolution describe different things. Specify the objective or SEM image setting, field width and length units used on your images.
- Manufacturers and models
- The Evident CX23 is an optical microscope, while the JEOL JCM-7000 NeoScope is a benchtop scanning electron microscope.
- What we check
- Compare the image scale against a dimensional reference at the settings where measurements are made. Record whether the result applies to the eyepiece, camera or saved image.
- Reference equipment and accuracy
- Use a calibrated stage micrometer for appropriate optical fields or a certified pitch artefact suitable for the SEM scale. Reference features and uncertainty must suit the image measurement.
- When to repeat the check
- Recheck after changing optics, camera settings or image calibration software. Keep the scale results for each configuration used for dimensional work.
Technical references Evident CX23 microscope instructions · JEOL instrument catalogue