Power sensor, frequency and connector
Specify the sensor model, frequency points, connector type and power range. For pulsed operation, distinguish peak power from average power and give the relevant pulse conditions. If a cable or attenuator is part of the measurement, define the reference plane.
A sensor factor stored for one frequency cannot simply be assumed valid at another. Identify the sensor factor source and selected frequency when comparing a source at two band edges. Otherwise a setup difference can appear to be a change in output power.
Specifications and service checks
A technician can zero a meter with the sensor fitted and later substitute another sensor on the same channel. Keep the sensor identifier beside each power reading. The mainframe, sensor and selected calibration factor jointly determine the indication.
- Range and configuration
- State the sensor frequency limits and level range in dBm or watts. A 0 dBm signal corresponds to 1 mW.
- Manufacturers and models
- Keysight N1914A is a dual-channel EPM power meter used with compatible sensors. Frequency coverage follows the fitted sensor.
- What we check
- Check the meter's electrical response and reference output as applicable, then compare the connected sensor and meter against known RF power at selected frequencies.
- Reference equipment and accuracy
- Use a calibrated reference sensor or power transfer system with known mismatch and connector performance. Distinguish a meter-only result from a sensor-and-meter result.
- When to repeat the check
- Reassess after connector damage or sensor overload and trend readings on the reference output.
Technical references N1914A EPM Series Dual-Channel Power Meter [Discontinued]