On this page
Two-wire and four-wire resistance measurements
Identify the working range, test current and two- or four-wire mode. If your bench compares successive parts using a zeroed fixture, specify the zero convention. An instrument-reading result and a result including the production fixture can then be distinguished.
A four-wire result across the contact tips of a fixture can exclude resistance that a two-wire check at the front panel includes. A clear record of those connection points helps explain a shift after a lead or contact has been replaced.
Specifications and service checks
For a winding checked immediately after a run, the resistance may be higher than when the winding is cold. Write down the temperature or cooling interval used by the workshop before attributing a difference between shifts to the low-resistance meter.
- Range and configuration
- A joint measuring 5 mΩ drops 50 mV at 10 A. Give both the expected resistance and test current so the reference can withstand the applied power and expose any self-heating error.
- Manufacturers and models
- Megger DLRO10 is a digital four-terminal low-resistance ohmmeter. A traditional Kelvin bridge uses a ratio comparison for similar low-resistance work.
- What we check
- Compare indicated values with calibrated four-terminal resistors; for a bridge, also check the ratio settings used. Current reversal can help quantify thermal offset.
- Reference equipment and accuracy
- Use low-resistance standards rated for the applied current, with separate current and potential connections. Include thermal voltage and self-heating in uncertainty.
- When to repeat the check
- Trend a stable low-ohm checkpoint. Recheck after lead, current-source or selector repairs.
Technical references DLRO10 and DLRO10X digital low resistance micro-ohmmeters