Milli ohmmeter calibration

At milliohm levels the contact and lead resistance can be a significant part of a two-wire reading. A four-wire connection senses voltage at the part separately. The test current matters too, particularly on a small component whose resistance changes as it warms.

Send an enquiry
Milli ohmmeter — equipment illustration
Equipment illustration
On this page

Two-wire and four-wire resistance measurements

Identify the working range, test current and two- or four-wire mode. If your bench compares successive parts using a zeroed fixture, specify the zero convention. An instrument-reading result and a result including the production fixture can then be distinguished.

A four-wire result across the contact tips of a fixture can exclude resistance that a two-wire check at the front panel includes. A clear record of those connection points helps explain a shift after a lead or contact has been replaced.

Specifications and service checks

For a winding checked immediately after a run, the resistance may be higher than when the winding is cold. Write down the temperature or cooling interval used by the workshop before attributing a difference between shifts to the low-resistance meter.

Range and configuration
A joint measuring 5 mΩ drops 50 mV at 10 A. Give both the expected resistance and test current so the reference can withstand the applied power and expose any self-heating error.
Manufacturers and models
Megger DLRO10 is a digital four-terminal low-resistance ohmmeter. A traditional Kelvin bridge uses a ratio comparison for similar low-resistance work.
What we check
Compare indicated values with calibrated four-terminal resistors; for a bridge, also check the ratio settings used. Current reversal can help quantify thermal offset.
Reference equipment and accuracy
Use low-resistance standards rated for the applied current, with separate current and potential connections. Include thermal voltage and self-heating in uncertainty.
When to repeat the check
Trend a stable low-ohm checkpoint. Recheck after lead, current-source or selector repairs.

Technical references DLRO10 and DLRO10X digital low resistance micro-ohmmeters