Copy the rated input and scale marks separately. If the meter is used on a live-zero loop, include the lower working value and the midrange indication used during checks. State the connection polarity and whether an external shunt is fitted.
On a panel scaled from 4 to 20 mA, a 12 mA input may represent halfway through a process range. Record whether the acceptance sheet asks for the electrical reading in mA or the process value printed on the dial.
Specifications and service checks
A low-current panel can be wired into either a passive input or a powered loop. Identify which arrangement supplies the current and whether the scale is centre-zero. This matters when the process normally swings from a small positive current through zero.
- Range and configuration
- For a 4–20 mA process loop the span is 16 mA and the midpoint is 12 mA. Choose points in the operating band as well as any portions of a wider dial that the plant uses.
- Manufacturers and models
- Rishabh's RISH VISTA range illustrates analogue panel instruments. Check whether the scale is direct-reading or set up for an external shunt or transformer.
- What we check
- Compare indication with applied current at several points, allowing an analogue pointer to settle. Include the shunt or transformer if the installed measuring chain is required.
- Reference equipment and accuracy
- Use a calibrated current source or a source measured through a calibrated shunt. Check leakage at microampere levels and shunt heating at higher current.
- When to repeat the check
- Track zero error and repeatability. After a shunt or movement change, verify the complete scale again.