On this page
AC and DC millivolt ranges
Provide the measured DC or AC range, polarity and labelled scale. For low DC levels, include the normal lead and connector arrangement. If the meter is part of a shunt assembly, identify whether the comparison covers the displayed current or the millivolt input alone.
A panel meter paired with a 75 mV shunt can be checked by applying a 75 mV reference signal to the reader. That establishes the panel response at the top of its scale. The shunt requires its own comparison to determine the voltage it produces at a given current.
Specifications and service checks
At the low end of a millivolt scale, warming a dissimilar-metal connection can produce a visible thermal offset. If the indication changes after someone handles a terminal, describe the connection materials and settling time as well as the applied voltage.
- Range and configuration
- One millivolt is 1,000 microvolts. Even small thermal voltages at dissimilar metal junctions can matter to a low-level result.
- Manufacturers and models
- Keysight 34420A is a low-level voltage and resistance meter. The lowest voltage ranges demand different connections from ordinary panel-voltage work.
- What we check
- Compare known positive and negative DC voltages at the selected points and evaluate the zero offset on the same range.
- Reference equipment and accuracy
- Use a calibrated low-voltage source or reference divider with low-thermal connections. Include offset stability and noise across the observation period.
- When to repeat the check
- Trend low-range zero and noise after moving the instrument or changing leads.
Technical references Advanced Benchtop Digital Multimeters: 34420A