Torque, speed and temperature functions
List the torque, speed or temperature functions you need checked. If you also use a reference material to check the whole system, add that as a separate item. A record identifying the fitted geometry and its serial number is more useful after an accessory replacement than a single result against the base unit.
For an instrument with interchangeable measuring systems, an order that names only the model does not identify the geometry used. Pair the fixture identification with its working gap or configuration and the temperature accessory. That combination helps explain why two runs on the same instrument may differ.
Specifications and service checks
If a measuring geometry is replaced, the instrument body may retain its calibration history while the new geometry has a new serial number and dimensions. Identify the fitted measuring pair and temperature accessory. A comparison can then address the configuration used in the reported result.
- Range and configuration
- The MCR 503 lists 0.15 nN·m minimum torque in oscillation and the MCR 503 Power lists 50 nN·m. Identify the particular drive and geometry before comparing a low-torque result.
- Manufacturers and models
- Anton Paar’s MCR family includes modular compact rheometers such as the MCR 503.
- What we check
- For a modular rheometer, identify the installed measuring geometry and temperature accessory before comparing results with a certified reference fluid. The calibration record should name that configuration.
- Reference equipment and accuracy
- Reference-fluid viscosity and uncertainty apply at specified temperatures. Traceable dimensional and temperature measurements matter when the gap or geometry affects the result.
- When to repeat the check
- Repeat relevant checks after geometry damage or bearing work. A stable reference-fluid trend helps choose the next service interval.
Technical references Anton Paar MCR rheometers · Anton Paar MCR torque specifications