Moisture balance calibration

A moisture balance reports mass loss while its heater runs a defined drying programme. Its mass indication, temperature control and calculation settings contribute different information. A weighing check alone does not verify the temperature reached under the sample pan.

Send an enquiry
Moisture balance — equipment illustration
Equipment illustration
On this page

Percentage loss on a wet or dry basis

On a wet-mass basis, a sample that falls from 10.00 g to 9.50 g has lost 5% of its starting mass. If the balance is set to express loss against the final dry mass, the same readings give about 5.26%. Copy the reporting basis, drying temperature, pan, programme and endpoint rule from the method. These let us distinguish a weighing or heater check from a performance check with your reference material.

Moisture balance weighing and heating

Give the drying temperature and endpoint criterion from the approved method. The weighing function, heating function and complete drying test answer different questions. Ask for each required check separately; an acceptable weight indication does not establish the temperature or the suitability of the sample's drying programme.

Specifications and service checks

A 1 g loss from a 10 g starting sample is 10% on the original wet-mass basis, or about 11.1% on the final 9 g dry-mass basis. Check which basis the display uses before comparing two moisture results.

Range and configuration
The OHAUS MB90 has a 90 g maximum sample load. Specify both the usual sample mass and the drying temperature used by the method.
Manufacturers and models
OHAUS MB90 combines a weighing cell and a halogen heater. Include the drying program when investigating an application result.
What we check
Verify mass indication using weights and check heater temperature with the maker’s temperature kit. Repeatability of moisture results is a separate application check.
Reference equipment and accuracy
Use calibrated weights and a compatible temperature reference with known uncertainty. A mass certificate by itself does not cover the heater.
When to repeat the check
Review both channels after a heater, pan support or load-cell change. Routine checks can identify which side needs attention between recalls.

Technical references OHAUS Moisture balance reference · OHAUS Moisture balance reference