Individual channel volumes and spacing
List the working volumes and tip system. For an instrument with adjustable spacing, describe the head arrangement used during the check. A per-channel report makes it possible to follow up one seal or cone without losing the history of the other channels.
In an eight-channel set, seven acceptable means and one high mean can still give an acceptable overall average. If the work needs a decision for each channel, request eight separately identified results at each chosen setting and ask for the applicable acceptance rule.
Specifications and service checks
A low delivery from channel eight can be lost in the total volume from all eight channels. Mark the orientation of the head and retain the order of individual channels when reviewing the results. This lets the operator locate the same channel on the bench afterwards.
- Range and configuration
- Eppendorf documents Xplorer heads with 8, 12, 16 or 24 fixed-spacing channels. The number of channels and the usable volume per channel are separate specifications.
- Manufacturers and models
- Eppendorf’s Xplorer and Xplorer plus are electronic piston pipettes. Gilson publishes model- and mode-specific data for PIPETMAN M.
- What we check
- Record each channel’s identifier in physical order and compare delivery at the selected volumes. A single under-delivering channel can disappear in a head-wide average.
- Reference equipment and accuracy
- Use a traceable gravimetric arrangement or another applicable validated volume method. Its uncertainty and resolution must support one channel’s smallest delivery.
- When to repeat the check
- Review after a head or seal change and after persistent channel-to-channel differences. Include the channel identity on interim checks.