On this page
Plate markings and the orifice drawing
For plate inspection, supply the plate marking and drawing with the dimensions you need recorded. For a complete flow measurement, include pipe bore, tapping arrangement, fluid conditions and transmitter range. Keep the dimensional result identifiable separately from the loop result.
Orifice plates and the requested result
Mark the dimensions that need results and retain the installation orientation. A dimensional report for the plate is separate from a calibrated discharge coefficient or an installed flow-system result. RTS measures the specified plate features and records the drawing revision and tolerance beside the results.
Specifications and service checks
After a pipe spool is reversed, the tag on an orifice plate may still agree with the instrument list. Check which face was upstream and how the plate was oriented relative to the taps. Inspection of the bore and edge can be tied to the actual installation.
- Range and configuration
- The diameter ratio is orifice bore divided by pipe inside diameter. Edge condition and the arrangement of pressure tappings also affect the inferred flow.
- Manufacturers and models
- WIKA FLC-OP plates are specified by bore, plate geometry and the application. Keep the original plate identification with its calculation sheet.
- What we check
- Measure the bore and inspect the inlet edge and flatness against the drawing. A flow-bench comparison addresses the discharge coefficient separately.
- Reference equipment and accuracy
- Use calibrated internal-diameter, dimensional and surface inspection equipment. Record the temperature at which dimensional measurements are made.
- When to repeat the check
- Inspect after polishing, erosion or fouling. Even a small edge change can matter before the next scheduled dimensional check.
Technical references Orifice plate, Orifice flange, Annular chamber - WIKA United Kingdom