Particle counter calibration

A counter on an equipment list may sample air or count particles in a liquid. Start with the medium and the units in the saved report. Counts per cubic metre of air and counts per millilitre of liquid belong to different measurement arrangements, even when the channel labels look alike.

Send an enquiry
Particle counter — equipment illustration
Equipment illustration
On this page

Size channels, sample volume and counting mode

Identify the size channels, sampling volume, inlet or flow cell and the counting mode. A cumulative result includes all particles above a stated threshold. A differential result refers to a band between thresholds. Give us the quantities and counting convention used on your routine report.

A file headed “100 particles” is incomplete if it omits the sample volume or channel. Fifty particles per millilitre and one hundred particles in two millilitres describe the same concentration if the counting conditions match. That arithmetic still says nothing about which particle sizes were counted.

Specifications and service checks

A counter that reports 120 particles in a sample and one that reports a concentration per cubic metre can both describe the same observation. Retain the sampled volume and channel threshold with the count. A displayed number by itself cannot establish agreement between the instruments.

Range and configuration
A liquid counter and an air counter can display the same units for particle diameter while covering quite different sizes. Particle Measuring Systems specifies 2–125 µm for the LiQuilaz II E20 liquid counter. The TSI 9306-V2 air counter has size channels spanning 0.3–10 µm.
Manufacturers and models
The LiQuilaz II E20 uses a liquid flow cell. The TSI AeroTrak 9306-V2 counts particles in air. Identify the sample medium before matching a counter to a service record.
What we check
Check the counter identification, the medium, its size-channel labels and the units used for sample volume. Retain the counting convention shown on the report so measurements from different counters can be compared.
Reference equipment and accuracy
Require traceability for the assigned particle-size reference and for the sample-volume measurement. State which channels the certificate covers.
When to repeat the check
Review the manufacturer’s service date together with optical and flow-service history. A changed optical head or pump merits review before routine use.