On this page
Frame, plates and loading weight
List the frame, plate size and loading weight supplied with the instrument. If temperature conditioning is also being checked, identify the oven or chamber separately. The load results belong to the perspirometer and its weights. The temperature results belong to the identified chamber.
Perspirometer loading assembly
RTS checks the specified dimensions and mass of the frame, plates and loading pieces, retaining their identities with the results. Associated conditioning equipment has its own temperature and timing measurements. The material-test record separately describes the chemical exposure and fabric assessment under the laboratory’s method.
Specifications and service checks
A worksheet that gives the applied force in newtons and the effective contact area in square metres allows the nominal pressure to be calculated in pascals. Include the loading assembly in the force value. Counting loose plates alone can miss part of the applied load.
- Range and configuration
- The loaded plate area and applied force determine specimen pressure. Quote the plate size and loading mass for the particular test method. James Heal’s fixture takes up to 20 test assemblies between separator plates.
- Manufacturers and models
- James Heal and SDL Atlas list perspiration testers using stacked specimen plates.
- What we check
- Inspect plate flatness and compare the loading assembly with a mass reference. Check a separate oven under its own temperature procedure if it forms part of the test.
- Reference equipment and accuracy
- Use calibrated masses or a suitable balance, a length reference for the plate area and a flatness reference sized for the plate.
- When to repeat the check
- Inspect plates for distortion before use. Recheck after replacing the load or clamping assembly.
Technical references James Heal product reference · SDL Atlas product reference · James Heal perspirometer specifications