Phototherapy unit testing

A phototherapy unit’s panel setting identifies an operating mode, while an irradiance meter measures the light reaching a defined plane. The distance from the light head and the sensor’s wavelength response belong in the engineering record alongside the reading.

Send an enquiry
Phototherapy unit — equipment illustration
Equipment illustration
On this page

Light-head position and irradiance readings

Identify the head or pad, lamp or LED configuration and the positions used for the equipment check. Record hours of operation if the unit displays them. A map of readings across the working area is more informative for follow-up than an unlabelled centre reading.

Irradiance values from different meters can vary with the wavelength band to which each detector responds. If you need a comparison with an earlier equipment report, include that meter’s detector reference, reported unit and measurement distance. That gives engineering a way to interpret the two records after a light head has been changed.

Phototherapy unit optical measurements

A reading from an ordinary illuminance meter is not interchangeable with a dedicated phototherapy measurement. Send the approved instrument and procedure requirements for review. No patient settings or treatment decisions belong in the service enquiry or proposed bench report.

Specifications and service checks

A centre reading alone cannot describe how irradiance varies across the working area. If a comparison across positions is needed, keep the detector type, position grid and distance the same when reviewing two service visits.

Range and configuration
The engineering measurement is spectral irradiance at the pad surface. Pad size, cover and instrument response influence the reading.
Manufacturers and models
GE HealthCare’s BiliSoft 2.0 is a fibre-optic phototherapy system using a light pad.
What we check
Use the manufacturer’s service arrangement and report the pad, measurement positions and instrument used for the comparison.
Reference equipment and accuracy
Choose a calibrated radiometer with the specified spectral band and response. A lux meter weights light for human vision and measures a different quantity.
When to repeat the check
Recheck after a pad, light source or optical assembly is changed. Keep the pad and controller identifiers together.

Technical references GE HealthCare product reference