PID controller calibration

A PID controller has an input indication, a set point and a control output. A calibration of the input can be carried out independently of how the installed process responds. Specify the input type and whether the output signal also enters your records.

Send an enquiry
PID controller — equipment illustration
Equipment illustration
On this page

Controller readings and process overshoot

If the maintenance issue is overshoot or slow recovery, describe the recorded process behaviour separately. Those symptoms are not, by themselves, a measure of input accuracy. Retain controller configuration and channel limits with the results for future replacements.

A controller can display the applied temperature correctly yet continue heating because the output action or relay wiring is wrong. If this is the reported fault, include the output and relevant operating logic in the enquiry. A display-only calibration would not resolve that part of the investigation.

Specifications and service checks

If a panel displays the right temperature but its output remains at a fixed percentage, identify whether the controller is in manual or automatic mode before requesting a test. An input-display comparison and a check of the commanded control output address two separate observations.

Range and configuration
The configured sensor type, displayed span and output type define the job. A relay contact rating in amperes does not describe temperature accuracy.
Manufacturers and models
Omron’s E5CC-T is a programmable temperature controller with configurable inputs and control modes.
What we check
For an indication check, simulate the configured input and compare the displayed process value. Separately identify any alarm or output that must be tested and record its configured action.
Reference equipment and accuracy
Choose a calibrated source for the configured thermocouple or RTD input. For a complete temperature loop use an independent thermometer at the measurement position.
When to repeat the check
Review after input configuration changes, controller replacement or drift in the process check. Preserve configuration details with the original result.

Technical references Omron E5CC-T specifications