On this page
Coupler and microphone adapter
Identify the pistonphone, coupler, adapters and the nominal frequency and level used. If a barometer reading is needed for the model’s correction, record its source. A result for one coupler configuration should remain identifiable when several microphone sizes share the device.
A sound level meter fitted to a pistonphone with a different microphone adapter may experience a different acoustic volume. Record which adapter was used on the day. A level marked on the case alone does not identify that coupling arrangement.
Specifications and service checks
Where the pistonphone manual specifies a correction for atmospheric pressure, note the barometer reading at use. This is especially helpful when comparing checks recorded at sites of different elevation with the same microphone and coupler.
- Range and configuration
- A pistonphone gives a fixed nominal sound pressure at a defined frequency and coupler volume. Atmospheric pressure and the fitted adapter may require corrections.
- Manufacturers and models
- GRAS 42AA is a pistonphone providing a nominal 114 dB at 250 Hz with its specified coupler.
- What we check
- Compare the acoustic output against a reference microphone in the proper coupler, recording frequency and the ambient conditions used for correction.
- Reference equipment and accuracy
- Use a traceable pressure microphone, calibrated readout and suitable barometric reference. Coupler fit and corrections belong in the uncertainty.
- When to repeat the check
- Check seals and output following transport or adapter damage. Review the regular interval against successive calibrated output levels.
Technical references GRAS 42AA pistonphone specifications