+91 70565 55582[email protected]Get a quoteRequest a call
RTS Calibration
Menu

Plain plug gauge calibration

A plain plug gauge checks a hole with a cylindrical working member. A double-ended limit gauge may have separate go and no-go diameters. Each end represents a different limit and can wear differently because one end passes through the work more often.

Send an enquiry
Plain plug gauge — equipment illustration
Equipment illustration
On this page

Markings on each plug end

Copy the size and tolerance marking from each end and identify any detachable members. If the working end is visibly polished near its nose, mark that region. A record by end and measuring position can show wear that a single diameter for the gauge would hide.

For a two-ended plug, stores may record one handle number while the individual members carry their own numbers. Include both sets of markings so a replacement member can be traced without retiring the handle history.

Specifications and service checks

A plug that enters a blind hole until trapped air cushions it may behave differently in an open ring. Identify the through or blind application and which member is being checked before a fit complaint is interpreted as size wear.

Range and configuration
RTS lists cylindrical plain plugs in the 5–25 mm band for the method on scope page 18. Record the individual GO and NO-GO design sizes, as well as the component nominal size.
Manufacturers and models
Vermont Gage supplies Class X plain pin/plug gauges. For a gauging pair record the maker’s marked class and the dimension of each member.
What we check
Measure diameter in several directions and at more than one height along the working land. Separate the results for the two ends and look for wear where parts first enter.
Reference equipment and accuracy
Use a calibrated length comparator and end standards or a suitable diameter measuring machine. Select the reference uncertainty for the gauge tolerance, which can be tighter than the part tolerance.
When to repeat the check
Look for entry wear and damage before using the gauge to release product. Recheck after replating or lapping.

Technical references Vermont Gage Plain cylindrical gauges · CC-4643 calibration scope, page 18

Enquire about Plain plug gauge

Service enquiry