On this page
Plunger travel and direction of approach
List the portion of the travel actually used and whether readings are taken on entering or returning from a high point. Recording both directions at common points can reveal a return difference that would be missed by a single run up the scale.
A cam inspection fixture may sweep a pointer up to a high spot and then read it as it returns. If that is how results are taken, readings made only as the plunger travels inward leave a practical question unanswered about the return direction.
Specifications and service checks
If the plunger is held at an angle to a moving slide, the motion along the spindle differs from the slide movement. Indicate the stem direction in the fixture drawing when comparing an installed dial reading with axis travel.
- Range and configuration
- A plunger gauge senses travel along its stem. On a model such as the Mitutoyo 2506S, identify the scale unit, revolution counter and total spindle stroke before choosing the points to check.
- Manufacturers and models
- Mitutoyo Series 2 includes standard plunger indicators in inch-reading versions. Identify the contact and back or stem mount supplied.
- What we check
- Move the plunger through known axial steps without side loading it. Record indication during advance and return. Check that the reading repeats after the tip is lifted and reseated.
- Reference equipment and accuracy
- Use a calibrated displacement tester with the gauge clamped by its approved mount. Allow for error caused by mounting pressure on the stem or misalignment of the spindle.
- When to repeat the check
- Inspect the plunger when the pointer hesitates or approaches the same point differently from opposite directions. A service or contact change justifies a fresh check.
Technical references Mitutoyo Dial Indicators