+91 70565 55582[email protected]Get a quoteRequest a call
RTS Calibration
Menu

Pneumatic pressure gauge calibration

A pneumatic gauge may be used on a regulated air line or as a portable check gauge. Two instruments with the same full scale can have very different working points if one checks supply pressure and the other monitors a low-pressure process.

Send an enquiry
Pneumatic pressure gauge — equipment illustration
Equipment illustration
On this page

Gas service and pressure reference

Which gas does the gauge measure, and what connection and pressure reference does it use? If another gauge on the plant disagrees, give us both settled readings and the regulator setting. We can include that pressure in the comparison.

A regulator may deliver a lower pressure when an air tool is running than when the outlet is closed. If that is when the gauge reads low, note the tool’s operating condition. It helps us distinguish a pressure drop under demand from an error on a steady gauge reading.

Specifications and service checks

A packing machine can momentarily dip below its air limit on each cylinder stroke. A dial check at steady pressure addresses the marked scale. A complaint about the depth of that short dip needs the approximate cycle time and where the dial sits relative to the cylinder.

Range and configuration
Gas-line gauges are specified by the maximum indicated pressure and dial divisions. Quote working points in the same pressure unit as the dial.
Manufacturers and models
WIKA 232.50 is an example for suitable gaseous media. Select the exact range and media compatibility on the order.
What we check
Compare successive increasing and decreasing readings using compatible gas pressure. The zero reading at atmospheric pressure is part of a gauge-pressure check.
Reference equipment and accuracy
Use a calibrated pneumatic reference or gas pressure balance. Account for reference uncertainty and any offset between pressure reference levels.
When to repeat the check
Review the interval after compressor pulsation, a pressure surge or a noticeable zero shift.

Technical references WIKA 232.50 and 233.50 Bourdon gauges

Enquire about Pneumatic pressure gauge

Service enquiry