+91 70565 55582[email protected]Get a quoteRequest a call
RTS Calibration
Menu

Positive displacement pipette calibration

A positive displacement pipette uses a piston that acts directly within the fitted capillary or tip. That accessory is part of the measuring mechanism. Record the capillary and piston type together with the pipette identity, particularly where several capacities fit the same handle.

Send an enquiry
Positive displacement pipette — equipment illustration
Equipment illustration
On this page

Delivery volumes with the capillary and piston

Give the delivery points required and the criterion used for that instrument and accessory combination. If replacement capillary-piston sets are part of the enquiry, identify how they are sampled. A result with one accessory size does not define deliveries from another capacity.

A pipette handle may accept two piston-and-capillary sizes with overlapping volume settings. Specify which consumable is fitted for the common setting. Two deliveries both labelled 50 µL can otherwise appear to represent the same instrument setup when different measuring assemblies were used.

Specifications and service checks

A handle fitted with a different capillary and piston combination can still carry the same asset sticker. For a delivery check, identify the consumable size fitted at the working setting. This ties the result to the parts that actually displace the liquid.

Range and configuration
Gilson lists 25 µL nominal for its MICROMAN E M25E with CP25 capillary piston. Other handles and capillary pistons have different usable ranges.
Manufacturers and models
Gilson MICROMAN E and its specified capillary pistons are an example of a positive-displacement system.
What we check
Measure the delivered volume with the handle and capillary piston identified. Repeated measurements at the selected settings separate bias from variation between deliveries.
Reference equipment and accuracy
Use an appropriately sensitive calibrated balance and suitable gravimetric conditions. The disposable piston is part of the measured arrangement.
When to repeat the check
Review delivery after changes to the piston type, handle repair or unexpected differences in a routine check. Keep consumable identity with the comparison.

Technical references Gilson MICROMAN E specifications · ISO 8655-6 gravimetric volume determination

Enquire about Positive displacement pipette

Service enquiry