Phase arrangement and clamp position
Identify the phase arrangement, voltage and current levels, frequency and power factor required. Note any direction marking on the jaws. If the normal instrument report includes active power and apparent power, specify both so the comparison covers the intended functions.
A clamp placed around one phase with voltage leads connected to another phase can produce a plausible-looking but incorrect power reading. Record the current channel and associated voltage pair, particularly when a three-phase survey uses one portable instrument sequentially.
Specifications and service checks
A welding line may draw large distorted currents for short intervals. A steady, sinusoidal comparison addresses a different operating case. Identify the averaging mode, waveform type and interval used when the clamp value is entered on a production sheet.
- Range and configuration
- A jaw current range in amperes is separate from the watt range. The selected wiring mode and phase relation determine the calculated power.
- Manufacturers and models
- HIOKI CM3286-50 combines clamp current measurement and voltage leads for power measurements.
- What we check
- Compare indicated power using controlled current and voltage with known phase relation. Record jaw orientation and include the supplied leads.
- Reference equipment and accuracy
- Use a traceable power standard and suitable current fixture. Check that any multi-turn fixture has a known turns count and supports the required phase accuracy.
- When to repeat the check
- Recheck after jaw damage or inconsistent readings when the conductor position changes.
Technical references AC CLAMP POWER METER CM3286-50