Precision balance calibration

A balance that weighs full packets near capacity may also be used to weigh small additions into a container. Give the display division and both working loads, including the container. These details show which measurements matter on your line.

Send an enquiry
Precision balance — equipment illustration
Equipment illustration
On this page

Working points with the tare vessel

List the weighing points and any tare vessel. If the operator uses the balance to detect a small difference between two items, state that difference. A sequence of repeated weighings at the working mass can help interpret a marginal sort or acceptance decision.

A bench using a large tare container for small ingredient additions may operate near mid-capacity even when the displayed net reading is low. Record the gross load on the pan as well as the requested net increment.

Specifications and service checks

A controlled check using an individual reference weight answers a different question from repeatedly dosing powder into a container. To investigate poor dispensing repeatability, keep the balance check and the sequence of dispensed additions as separate observations.

Range and configuration
The OHAUS PR2202 has a 2,200 g capacity and 0.01 g readability. When a container is tared, both container and sample still use part of the available weighing capacity.
Manufacturers and models
OHAUS PR2202 is a precision-balance example. A different suffix can change capacity or readability.
What we check
Compare working loads with calibrated weights, repeat a representative load and check position sensitivity on the pan.
Reference equipment and accuracy
Choose weight class and certificate uncertainty against the smallest permitted weighing error. A class marking does not replace the individual weight certificate.
When to repeat the check
Set the recall using check-weight performance at both the light and heavy end. Review it after a change of bench or weighing range.

Technical references OHAUS PR Series Precision

Enquire about Precision balance

Precision scale

If stores calls the instrument a precision scale, copy the actual display division and capacity from the label. The word precision alone does not distinguish between a fine bench balance and a higher-capacity platform instrument.

A scale may be used in grams on the bench and kilograms on its external display. Compare both against the same unchanged load when printed entries differ by a factor of 1,000, and retain the displayed unit with the value.

Range and configuration
An OHAUS PR2202, for example, weighs to 2,200 g with a 0.01 g display division. The smallest sample a process can accept depends on measured uncertainty.