Accuracy specification and reference pressure
Give the maker’s accuracy specification, unit, range and reference type. Include the pressures at which you actually use it as a reference and the smallest dial division or display step. If acceptance is required, identify which revision of the specification applies.
If a maintenance team transfers corrections from a reference gauge to its working sheets, identify the convention on the certificate. A reference reading high by one division requires a correction of the opposite sign when estimating the applied pressure.
Specifications and service checks
Think about a reference dial shared between two benches with different limits. A tolerance expressed as 0.1% of a 100 bar span allows 0.1 bar at a 5 bar working point. Work out whether that allowance suits the low-pressure bench before choosing this gauge as its reference.
- Range and configuration
- The WIKA CPG1500 series includes gauge-pressure ranges up to 1,000 bar and different accuracy options. Confirm the individual full scale and whether the permitted error is specified as a percentage of span or reading.
- Manufacturers and models
- WIKA CPG1500 is a digital reference gauge available in several accuracy options. Mechanical test gauges have a dial class instead.
- What we check
- Compare the working points on increasing and decreasing pressure and record the display unit and zero setting. Retain the as-found readings before any correction.
- Reference equipment and accuracy
- For a precision gauge used as a master, compare its permitted error with the uncertainty achievable at each set point. Include reference stability, pressure head and the digital gauge's own repeatability.
- When to repeat the check
- For a gauge used as a master, review the calibration history and intermediate comparisons with an independent check gauge.