On this page
Working fluid and reference connection
Identify the comparator working fluid, connection arrangement and required pressure span. If the order includes a reference gauge or digital module mounted on the comparator, list it separately. Pressure generation, leak performance and the reference indication are distinct parts of the assembly.
If a test dial is removed and refitted on a different adapter, the effective elevation and trapped volume may change. Record those changes when investigating a discrepancy between two runs made with the same comparator and reference.
Specifications and service checks
A comparison stand shared by coarse production dials and a low-range reference may need fine movement near one particular set point. Tell us whether the problem is reaching that point or holding it after the adjustment stops. Those observations concern different parts of the comparison setup.
- Range and configuration
- Select the comparator by maximum pressure, medium and usable adjustment range. A nominal 1,000 bar comparator is not itself a 1,000 bar pressure standard.
- Manufacturers and models
- WIKA CPP1000-H is a hydraulic comparison test pump. A separate pressure reference is fitted for a comparison.
- What we check
- Compare the gauge under test and the reference at a common pressure. Check the comparator for leakage and establish the level of the two measuring ports.
- Reference equipment and accuracy
- The master is the calibrated gauge or pressure balance. The comparator provides the shared pressure, with a head correction where reference levels differ.
- When to repeat the check
- Inspect seals and ports between jobs. Schedule the reference separately and review the comparator after leaks or maintenance.
Technical references WIKA CPP comparison pump specifications