Input scaling on the pressure display
Give the input type, scaling and display unit. For a test of the full indication, identify the matching sensor and cable. Note any zero suppression or offset so a technician can reproduce the readings the operator sees.
A receiver scaled from 4–20 mA to 0–16 bar should display 8 bar at a 12 mA input if it uses a linear relationship. That check exercises the configured indication. Applying 8 bar at the transmitter also includes its sensor and current output.
Specifications and service checks
When a front-panel pressure display has an adjustable decimal point, a setting change can make the correct input appear ten times larger. Provide the previous and present unit labels with a known input condition if the issue followed replacement of the panel unit.
- Range and configuration
- Pressure units on the screen can change without changing sensor span. List the range and pressure mode for each connected input.
- Manufacturers and models
- WIKA CPG2500 is a precision pressure indicator that can be fitted with different transducers.
- What we check
- For the instrument as used, compare a known pressure at the transducer with the displayed value. A separate simulation can isolate a faulty electrical input.
- Reference equipment and accuracy
- A calibrated reference controller or pressure balance establishes the pressure. Keep any external sensor serial number with its channel.
- When to repeat the check
- Review each channel's history independently. Replacing a remote transducer changes the measuring chain.
Technical references Precision Pressure Indicator CPG2500