Matching pressure to the recorded trace
Include the recorder channel, pressure sensor, recording interval and report file or chart scale. If you need agreement between the applied pressure and an exported value, specify the exported field. For a chart unit, identify both the pressure division and chart speed.
If a plant trend stores one average per minute while the recorder displays instantaneous pressure, a short excursion can be visible locally and smaller in the saved record. Keep the averaging and storage settings with any comparison between the display and the downloaded trace.
Specifications and service checks
A batch trace that ends just before the pressure peak may reflect a recording trigger or stop condition. Include the trigger setting and the stored sample interval. A steady-pressure comparison by itself gives little information about a recorder missing a brief event.
- Range and configuration
- A chart recorder has a pressure scale and a chart period. Verify the fitted chart range and whether a complete revolution represents hours or days.
- Manufacturers and models
- Barton 202E is an example of a mechanical differential-pressure chart recorder. Electronic recorders also need their exact input and timebase identified.
- What we check
- Compare the position or recorded value against applied pressures. Where timestamps matter, check elapsed time or chart advance separately.
- Reference equipment and accuracy
- Use a calibrated pressure standard plus a traceable time reference for a timebase check. Pen width or logging interval can limit resolution.
- When to repeat the check
- Review when pen linkage, chart movement, sensor or clock is serviced, even if the routine calibration remains in date.
Technical references Sensia BARTON 202E chart recorder