On this page
Matching the wet and dry bulbs to the humidity reading
A useful order might read ‘sling psychrometer, dry bulb and wet bulb as a pair, humidity read from the attached chart’. An aspirated digital instrument may instead calculate humidity internally. For either type, the two thermometer readings are valuable when a comparison disagrees. With calculated results, include the chart or pressure setting used on the day.
Psychrometers
Retain the wick and water requirements from the operating procedure as part of the configuration record. Calibration of the two temperature indications and evaluation of the resulting humidity are different tasks. Ask RTS to review which results can be provided, with the conditions and calculation clearly stated.
Specifications and service checks
A dry-bulb reading alone cannot reproduce the humidity value calculated from a wet- and dry-bulb pair. Retain both temperatures at the same instant along with the pressure or table convention used. Those details allow a reviewer to reconstruct the displayed humidity.
- Range and configuration
- Brannan specifies −5 to +50 °C with 0.5 °C divisions for the thermometers in its cited whirling hygrometer. The calculated humidity depends on the pair of readings and the applicable conversion.
- Manufacturers and models
- Brannan supplies a whirling hygrometer with a paired wet- and dry-bulb arrangement.
- What we check
- Compare the wet- and dry-bulb thermometers individually. Check which table or calculation the instrument uses when a humidity result is requested.
- Reference equipment and accuracy
- Use a traceable temperature reference for the bulbs. For an assembled humidity comparison use a reference hygrometer and account for the pressure term.
- When to repeat the check
- An altered wick, broken bulb or disagreement between the two dry thermometers warrants attention before the normal review date.
Technical references Brannan whirling hygrometer · NIST humidity measurement services