On this page
Pulse width, repetition rate and termination
Give the pulse amplitude, termination, polarity, width and repetition rate. Identify whether timing is measured from a trigger edge or between two channel outputs. For a critical edge measurement, include the threshold or convention used to define the edge.
A 1 kHz pulse train has a period of 1 ms. At 10% duty cycle, the nominal on-time is 0.1 ms. A frequency reading alone can look correct even when the pulse stays high too long. Record width as well as repetition rate where both affect the test.
Specifications and service checks
A fast edge at the generator output can look slower after a cable and fixture are attached. Identify where the time measurement is made and what termination is fitted. Comparing pulse width at one node with rise time at another can hide the original complaint.
- Range and configuration
- For a 1 kHz train, the period is 1 ms; a 10% duty cycle gives a nominal 100 µs high time. Rise time is a separate specification.
- Manufacturers and models
- Keysight 81160A is a pulse/function generator. Specify the timing and termination used for the output under test.
- What we check
- Compare high and low levels, period and width with the stipulated load. Measure edge times at the threshold percentages given for that model.
- Reference equipment and accuracy
- Use a calibrated high-bandwidth oscilloscope or sampling system, suitable termination and traceable timebase. Reference rise time must be small enough to resolve the edge.
- When to repeat the check
- Review after output damage or a change in edge behaviour at a fixed test condition.
Technical references Pulse Generators: 81160A