Pycnometer calibration

A pycnometer establishes density from the mass of liquid occupying a defined volume at a specified temperature. The stopper or cap helps define that volume and should remain with the body to which it belongs.

Send an enquiry
Pycnometer — equipment illustration
Equipment illustration
On this page

Matching vessel and stopper at temperature

Send the nominal capacity, operating temperature and both component markings. Specify whether the result required is the vessel volume or a check of a complete density measurement including your balance. For multi-part sets, identify the thermometer or capillary fitted during the comparison.

If a bottle and stopper are weighed dry and then filled, their mass difference represents the liquid in the defined volume. Substituting a stopper between the two observations invalidates that pairing even when both stoppers share the same nominal size.

Specifications and service checks

For an illustrative density calculation, a liquid adds 24.960 g in a bottle with an assigned volume of 25.000 mL at the measurement temperature. The quotient is 0.99840 g/mL. Using a nominal “25 mL” bottle size instead of its assigned volume can change the last digits.

Range and configuration
The marked capacity is nominal. Density calculations use the actual volume at a reference temperature, or a ratio of sample and water masses under stated conditions.
Manufacturers and models
DWK's MBL Gay-Lussac density bottle has a fitted capillary stopper. The stopper forms part of the calibrated volume.
What we check
Determine the mass of water contained with the fitted stopper at known temperature. Report the volume and identify the stopper.
Reference equipment and accuracy
Use a calibrated balance, water of known density and a calibrated thermometer. Include filling repeatability and evaporation where relevant.
When to repeat the check
Repeat the volume determination after replacement of a stopper or alteration of the capillary.

Technical references MBL® Density Bottle adjusted

Enquire about Pycnometer

Specific gravity bottle

If a stopper has been replaced, identify it before using the bottle’s previous capacity value. The fill level and volume can change with a different stopper even when both pieces fit.

A matched bottle can be used for a liquid-to-water comparison. If the liquid adds 21.60 g and water adds 24.00 g in the same filled volume at the same temperature, their mass ratio is 0.9000. The ratio has no unit. Keep the matching stopper with the bottle for both weighings.

What we check
Use the matching stopper and determine the bottle's water capacity at controlled temperature. For crossed bottles or stoppers, re-establish the volume of the actual pair.