Reference thermometer calibration

The same reference thermometer might be used near 0 °C for one job and near 80 °C for another. Your uncertainty requirement can be different at the two points. Tell us the temperatures where the reference is used and whether you apply the correction on the probe certificate or read a corrected value from its paired indicator.

Send an enquiry
Reference thermometer — equipment illustration
Equipment illustration
On this page

Reading a correction from the certificate

Where the laboratory uses corrections from an earlier certificate, include the correction history and sign convention. Compare the correction at the same temperature on successive certificates to see where the reference may have drifted.

When a reference comes back with a fresh certificate, check how the result is labelled. With a reference of 50.00 °C and an indication of 50.03 °C, the indication error is +0.03 °C and the correction to add is −0.03 °C. Use the sign convention on your own certificate when entering a correction in the worksheet.

Specifications and service checks

A reference probe may travel between two precision readouts. Mark which readout holds its calibration coefficients and which input channel is normally used. If the probe goes out by itself, retain enough information to re-establish the correct configuration when it returns.

Range and configuration
The Fluke 1523 has one measuring input and the 1524 has two. A probe’s calibrated span and fitted coefficients define the actual temperature coverage of a reference system.
Manufacturers and models
Fluke Calibration offers 1523 and 1524 reference readers and dedicated platinum probes for different ranges.
What we check
Check the probe-and-readout combination against a higher-order reference. If separate certificates are used, keep the serial number, channel and coefficients together.
Reference equipment and accuracy
The comparison reference and stable source must leave enough uncertainty allowance for the work this reference will later certify. Resolution alone is a poor selection criterion.
When to repeat the check
Track reference checks between calibrations. Drift at a check point, shock or a coefficient change should prompt review of the instrument and recent results.

Technical references Fluke 1523 and 1524 guide · BIPM guide to industrial platinum resistance thermometers · Fluke Calibration reference thermometer catalogue