Refractometer testing

A refractometer may display refractive index directly or express it on a product scale such as Brix. Temperature and the selected scale both matter. Automatic compensation can make the displayed result different from a raw reading at the measured sample temperature.

Send an enquiry
Refractometer — equipment illustration
Equipment illustration
On this page

Brix and refractive-index readings

Does the operator release the batch on a Brix reading or a refractive-index reading? A reference liquid whose certificate gives refractive index at a stated temperature needs to be interpreted on the correct scale. Send the working range, the temperature setting and any compensation selected. For an installed unit, name the sensor and display. For a handheld unit, identify the prism and the references already used on site.

Refractometer scale and temperature

Record the selected scale, temperature control or compensation settings, and the reference material used. Include its certificate, expiry and measurement temperature where specified. A comparison using one reference liquid does not validate a concentration conversion for an unrelated mixture; the analytical method must identify that separate relationship.

Specifications and service checks

A sugar solution and a different liquid can give similar refractive indices. A scale marked °Brix interprets the optical reading for a particular application. Specify the liquid and whether the release decision uses refractive index or the instrument’s converted value.

Range and configuration
The ATAGO PAL-1 reads 0.0–53.0% Brix. An instrument marked in refractive index or on a coolant-specific scale calls for a reference certified in the corresponding quantity and at the stated temperature.
Manufacturers and models
The ATAGO PAL-1 is a digital Brix model. The MASTER-53α is a hand-held optical model in the same nominal Brix range, with a boundary line read through the eyepiece.
What we check
Check the zero and compare readings against appropriate certified solutions across the working scale. Record the reference temperature and whether the instrument’s automatic temperature compensation was active.
Reference equipment and accuracy
Use certified sucrose mass-fraction standards for the Brix scale, or certified refractive-index liquids for an index scale. Include a calibrated thermometer when temperature is measured independently.
When to repeat the check
A water zero is a useful routine check. It cannot show scale error at a production concentration. Review drift using a reference near that concentration and recheck after damage to the prism.

Technical references ATAGO Refractometer PAL-1 · ATAGO Hand-Held Refractometer