Refrigerated centrifuge calibration

A refrigerated centrifuge reports temperature as well as rotational speed. The indicated temperature can refer to a particular sensor in the chamber, so the location of a reference reading matters. Identify the chamber, rotor and instrument temperature channel on the job sheet.

Send an enquiry
Refrigerated centrifuge — equipment illustration
Equipment illustration
On this page

Speed, time and temperature at the working position

State whether the requested comparison concerns the displayed temperature, conditions at an identified position, speed or elapsed time. The report should preserve the operating configuration for each reading, especially when temperature checks and speed checks are performed under different conditions.

A service sheet may show agreement at the chamber probe, while a separate logger in a holder records another temperature. Before combining the figures, record where and under what operating condition each was obtained. They can be valid observations of different parts of the instrument.

Specifications and service checks

A department recording both speed and temperature should identify where the temperature is sensed. On models with a chamber sensor, the cabinet indication and the temperature at a rotating sample position are distinct measurements. Define which observation the equipment review is intended to support.

Range and configuration
The drive and the fitted rotor have separate speed limits. Indicated rpm and relative centrifugal force are different quantities. Refrigerated models also have a separate temperature channel.
Manufacturers and models
Eppendorf’s 5804 and 5804 R are benchtop examples. Beckman Coulter’s Avanti family includes floor-standing systems with rotor-specific limits.
What we check
Request the speed comparison for the identified rotor and any separate comparison for temperature. Define the temperature measurement location and the manufacturer-approved access used for the equipment.
Reference equipment and accuracy
A calibrated tachometer or authorised service reference establishes speed. A time reference and calibrated thermometer are needed if those quantities are included.
When to repeat the check
Review after a drive or speed sensor repair. Rotor condition and lifetime inspections belong in the maintenance record alongside speed calibration.

Technical references Eppendorf 5804 and 5810 series manual · Beckman Coulter Avanti rotor catalogue · Stanhope-Seta photo tachometer