Resistivity meter calibration

A resistivity meter for water may share its cell with a conductivity measurement, but the units and displayed range differ. Identify the cell constant, temperature channel and whether the reported result is temperature compensated. Copy the units exactly from the process display.

Send an enquiry
Resistivity meter — equipment illustration
Equipment illustration
On this page

Meter input and the paired measuring cell

For an electronic check, record the input applied to the meter and the configured cell data. For a check of the meter and cell together, retain the cell identifier with the result. This allows a replacement cell to be traced later.

For the same water at the same temperature and on the same compensation basis, 1 µS/cm corresponds to 1 MΩ·cm. That reciprocal relationship can help spot a unit or reporting mistake. Comparing an uncompensated result with a value corrected to a reference temperature would not be equivalent.

Specifications and service checks

A water system may display resistivity in MΩ·cm and use conductivity in µS/cm for its trend. At the same temperature basis, 2 MΩ·cm corresponds to 0.5 µS/cm. Check the units and compensation settings before treating the two numbers as a disagreement.

Range and configuration
A conductivity/resistivity channel on the cited Thornton M300 lists 5,000 Ω·cm to 500 MΩ·cm for one configuration. Choose the actual cell and reported compensation mode before assigning a working span.
Manufacturers and models
METTLER TOLEDO Thornton M300 and Hanna DiST testers illustrate different calculations based on conductivity.
What we check
Compare the applicable input or the full sensor chain with a traceable reference. For a TDS display, record the factor used to convert conductivity to its indicated TDS value.
Reference equipment and accuracy
Use suitable conductivity standards or a traceable electrical simulation for the electronic channel. For high-resistivity water measurements, contamination can dominate a liquid comparison.
When to repeat the check
Recheck after cell changes and after changes to the temperature compensation or TDS conversion factor. Trend results near the normal water-quality limit.

Technical references METTLER TOLEDO Thornton M300 specifications · Hanna DiST tester documentation