On this page
Meter input and the paired measuring cell
For an electronic check, record the input applied to the meter and the configured cell data. For a check of the meter and cell together, retain the cell identifier with the result. This allows a replacement cell to be traced later.
For the same water at the same temperature and on the same compensation basis, 1 µS/cm corresponds to 1 MΩ·cm. That reciprocal relationship can help spot a unit or reporting mistake. Comparing an uncompensated result with a value corrected to a reference temperature would not be equivalent.
Specifications and service checks
A water system may display resistivity in MΩ·cm and use conductivity in µS/cm for its trend. At the same temperature basis, 2 MΩ·cm corresponds to 0.5 µS/cm. Check the units and compensation settings before treating the two numbers as a disagreement.
- Range and configuration
- A conductivity/resistivity channel on the cited Thornton M300 lists 5,000 Ω·cm to 500 MΩ·cm for one configuration. Choose the actual cell and reported compensation mode before assigning a working span.
- Manufacturers and models
- METTLER TOLEDO Thornton M300 and Hanna DiST testers illustrate different calculations based on conductivity.
- What we check
- Compare the applicable input or the full sensor chain with a traceable reference. For a TDS display, record the factor used to convert conductivity to its indicated TDS value.
- Reference equipment and accuracy
- Use suitable conductivity standards or a traceable electrical simulation for the electronic channel. For high-resistivity water measurements, contamination can dominate a liquid comparison.
- When to repeat the check
- Recheck after cell changes and after changes to the temperature compensation or TDS conversion factor. Trend results near the normal water-quality limit.
Technical references METTLER TOLEDO Thornton M300 specifications · Hanna DiST tester documentation