Full scale designation and indenter

For each range you use, give the full Rockwell scale designation, the fitted indenter and the anvil. State whether the procedure uses regular or superficial scales. Keep the reference block number and scale beside its reading, particularly if more than one block is used on the same machine.

A machine switched from a ball to a diamond indenter between jobs should have its readings linked to the indenter used at the time. The scale designation and block serial number on each entry make that change visible during a later audit.

Specifications and service checks

An entry such as “60 Rockwell” leaves the scale open. Keep the letter designation with the reading and the block assignment, especially on a machine switched between scales during the same shift.

Range and configuration
Wilson 574R regular-scale machines use total forces of 60, 100 and 150 kgf. Superficial models use 15, 30 and 45 kgf. Identify the actual scale, such as HRC or HR15N.
Manufacturers and models
Buehler Wilson 574R, 574S and 574T differ in the scales they support.
What we check
Use certified blocks appropriate to the selected scale and inspect repeatability across properly spaced impressions. Direct checks address test forces, depth measurement and test cycle as applicable.
Reference equipment and accuracy
Use scale-specific reference blocks and suitable standards for force and depth. A block value certified on one Rockwell scale cannot establish accuracy on another scale by conversion.
When to repeat the check
Check the working block after an indenter or anvil change and at the frequency specified by the test procedure. Investigate a persistent shift before measuring new work.