On this page
Scale, contact time and support
Show the scale, stand and contact arrangement, and specify the time at which the operator reads the result if the method defines one. A request for a check on reference material should identify that material and its assigned scale.
An incoming rubber inspection may specify a reading at a set time after contact. If a bench instrument stores a later reading automatically, identify the time actually used for acceptance. An instant display reading may differ even on the same specimen.
Specifications and service checks
A thin gasket on a rigid bench can give a different reading from a sufficiently thick sample of the same compound. Note specimen thickness and support alongside the durometer type if the discrepancy appears only on finished parts.
- Range and configuration
- Shore A and Shore D instruments each have a 0–100 scale but use different indenters and spring forces. The RTS scope lists both of these scales.
- Manufacturers and models
- Bareiss HPE III Basic is available in Shore configurations. Record the letter on the instrument.
- What we check
- Compare Shore indication with known indentation depths over the working scale. Check the foot, indenter and prescribed spring forces, and identify a bench stand or added mass if the operator uses one.
- Reference equipment and accuracy
- Use a traceable displacement tester and force reference suited to the selected Shore scale. Check blocks help monitor performance between formal verifications.
- When to repeat the check
- Inspect the indenter before a test series. Dropping the unit or replacing a spring warrants reassessment on the affected scale.
Technical references Bareiss HPE III BASIC · CC-4643 calibration scope, page 22